H1 AOS
daniel.vander-hyde@LIGO.ORG - posted 13:04, Tuesday 10 July 2018 (42839)
ITMY Hartmann sensor angular sensitivity to SR3 and ITMY

TVo, Stefan, and Danny

We adjusted the pitch and yaw of ITMY and SR3 by a microradian while observing the x and y prism channels as well as the Hartmann gradient/contour plots. We also looked at the spherical and cylindrical power but didn't see any significant change.

ITMY Pitch (1 µrad) Yaw (1 µrad)
Prism X  <0.1 µrad  1.75 µrad
Prism Y 1.8 µrad  <0.1 µrad

 

SR3 Pitch (1 µrad) Yaw (1 µrad)
Prism X   <0.1 µrad 4 µrad
Prism Y  4.3 µrad  <0.1 µrad

We noticed that there is not a lot of cross coupling between Pitch to Prism X and Yaw to Prism Y. 

Note that there is a fringing pattern in the contour plots (see attached) as we moved both ITMY and SR3.

Images attached to this report