X1 SUS
mark.barton@LIGO.ORG - posted 10:06, Monday 24 September 2012 - last comment - 12:16, Tuesday 25 September 2012(4278)
Resuming BSFM02 testing
Mark B.

Picking up where I left off on Friday, aiming towards getting TFs.
Comments related to this report
mark.barton@LIGO.ORG - 09:48, Tuesday 25 September 2012 (4287)
Mark B.

Done for the day. Had chronic problems trying to get TF scripts to run, so reverted to a free-swing test to at least get the mode frequencies. Most of the LF modes seem to be where they're supposed to be but there's a huge anomalous peak at 11 Hz in all DOFs but especially V and R.
Non-image files attached to this comment
travis.sadecki@LIGO.ORG - 12:16, Tuesday 25 September 2012 (4290)

I went in to diagnose what might be causing the anomalous 11Hz peak and found that the Top Stage blade EQ stop (aka the goalpost) was locking the right side blade down.  I released this, then made a few small adjustments to correct for the now fully suspended chain/BOSEM alignment.