Reports until 14:06, Wednesday 18 July 2018
H1 AOS
craig.cahillane@LIGO.ORG - posted 14:06, Wednesday 18 July 2018 - last comment - 18:06, Wednesday 18 July 2018(42957)
ALS X PDH and PLL Spectra
I went down to X end and repeated the ALS measurements done for Y end.

ALS X green visibility is 30%.  Free swinging PDH peaks give ~750 mVpp demodulated reflection response.  
With a cavity pole of 275 Hz, we roughly calibrated the PDH error signal with 2*cavityPole/PDHvolts = 733 Hz/V.

Plot 1
PLL UGF is 27 kHz
PDH UGH is 8 kHz

Plot 2
PLL error spectrum

Plot 3
PDH error spectrum

Plot 4
Estimated transmission frequency noise from PDH error spectrum with arm pole applied.
In loop ALS X Transmission Frequency Noise RMS is 1 Hz.

Now that ALS COMM is locking, we will get an out-of-loop measurement of ALS frequency noise soon.
Images attached to this report
Comments related to this report
craig.cahillane@LIGO.ORG - 18:06, Wednesday 18 July 2018 (42963)
Took a quick look at ALS COMM PLL output to estimate what kind of frequency noise we're seeing.

The channel H1:ALS-C_COMM_PLL_CTRL_OUT_DQ is already calibrated into micrometers.  I uncalibrated it back to Hz for the plot below.

ALS COMM RMS = 25.8 Hz = 183 pm
Images attached to this comment
Non-image files attached to this comment