I went down to X end and repeated the ALS measurements done for Y end.
ALS X green visibility is 30%. Free swinging PDH peaks give ~750 mVpp demodulated reflection response.
With a cavity pole of 275 Hz, we roughly calibrated the PDH error signal with 2*cavityPole/PDHvolts = 733 Hz/V.
Plot 1
PLL UGF is 27 kHz
PDH UGH is 8 kHz
Plot 2
PLL error spectrum
Plot 3
PDH error spectrum
Plot 4
Estimated transmission frequency noise from PDH error spectrum with arm pole applied.
In loop ALS X Transmission Frequency Noise RMS is 1 Hz.
Now that ALS COMM is locking, we will get an out-of-loop measurement of ALS frequency noise soon.