I did another cavity scan for the OPO in vacuum.
The image is attached as below. (The y-axes units are in volts. )
- The 00 mode for green calculated from transmitted power is now ~83% . This is higher than the last measurement here.
- The 20 mode being 13% of the 00 mode, which is also better.
- The PZT calibration is still ~20V/FSR.
Next steps:
- OPO Tec Servo controller setting for vacuum. Currently the servo works, but when we set a new temperature, the temperature goes to the other direction and then comes back.
- OPO loops improvements. (Nutseinee and Daniel are working on)
- Set the position of the crystal to the optimized location.
- Send SEED in, and measure the SEED beam mode for homodyne path mode matching, with the temperature for dual resonance without NLG.