Reports until 13:15, Tuesday 14 August 2018
H1 TCS (AWC, TCS)
daniel.vander-hyde@LIGO.ORG - posted 13:15, Tuesday 14 August 2018 - last comment - 14:57, Wednesday 15 August 2018(43420)
Movement of HWS ITM periscope mirrors

TVo, Danny

To get a better idea of how the picomotor movements effect the movement of the Hartmann beam across the ITM, we decided to move them by a larger amount. To make sure there isn't any clipping I went out to the HWS table and removed the Hartmann plate to view the beam on the ccd while moving. 

The following moves were made: 

ITMX upper periscope mirror: 700 counts to the right

ITMX lower periscope mirror: 500 counts to the right

ITMY upper periscope mirror: 500 counts to the left. 

Attached are the current picomotor settings.

A ring heater test will be coming soon. 

Images attached to this report
Comments related to this report
daniel.vander-hyde@LIGO.ORG - 14:57, Wednesday 15 August 2018 (43450)AWC, TCS

TVo, Georgia, Danny

After a discussion at the commissioning meeting about using CO2X for a contrast defect measurement, we moved the Hartmann beam for ITMX back to where it was before this comment in order to make an attempt at centering CO2X sometime tomorrow. The plan is to do a contrast defect measurement and have CO2X (about) centered sometime before Friday so we can use CO2X to improve it on Friday.