Edgard, Hang.
On Tuesday, we designed and installed the ISIFF filters for ITMX. The procedure followed was similar to the one described in LHO:42875.
The attachment shows the filter installed in the ISIFF_L2P filter bank. The data points correspond to (ISI-L --> L3OPLEV_P)/(M0-P --> L3OPLEV_P), while the orange trace was fitted using IIRational.
The fitting code only used data points with a coherence >0.5, which was set as an arbitrary cutoff. The fit is reasonably good up to ~1 Hz.
More details on the performance of the filter, as well as information on the script used for the fitting will be posted soon.
_______________________________________________________________________________________________________________