Reports until 13:14, Friday 17 August 2018
H1 TCS (TCS)
georgia.mansell@LIGO.ORG - posted 13:14, Friday 17 August 2018 - last comment - 13:27, Sunday 19 August 2018(43492)
CO2X cenetering on ITMX

Danny TVo Georgia

This morning we pico'd the top periscope mirror on the CO2X table, it is now roughly centered on ITMX. From last night's long-duration low-power ring heater test we know where the center of the test mass is relative to the centre of the HWS beam. We then used 5-10 minute blasts of 1-2 W CO2 to see where the CO2 beam is, and approximately aligned this with the centroid of the ring heater contours.

 

The contours from the ring heater test are shown in the first attachment, the center is roughly (+12.5, +12.5 mm) relative to the Hartmann center. 

The second attachment shows the CO2 central heating, and the third shows the annular heating. NOTE: we moved the picomotor -700 counts (~4mm right (very rough) on the Hartmann camera) after taking these screenshots and didn't get a chance to take another CO2 test. The final position of the top periscope mirror is -11787,-1500 counts.

We removed the iris before these images were taken and are not sure why we don't see the full annular mask, particularly in yaw. (We also no longer seem to be limited by the confusing ghost beam which we put the iris in to remove).

The fourth attachment shows the FLIR camera image of the CO2 beam on the table with the central mask in - the beam is not quite centered on the mask in yaw which is something we might want to fix in the future.

Images attached to this report
Comments related to this report
daniel.vander-hyde@LIGO.ORG - 11:52, Sunday 19 August 2018 (43515)AWC, TCS

Some more information on the ITMX RH test from this post. The trend data is attached and I've posted a couple of the contour plots showing displaying two different times when ITMX lensing reached a steady state (as suggested by the HWS spherical power). 

Images attached to this comment
georgia.mansell@LIGO.ORG - 13:27, Sunday 19 August 2018 (43517)

Additional additional info: we retook the annular mask CO2 test on friday evening. It looks like we were better centered *before* the final 700 count move and should maybe consider moving back. Two screenshots from different stages of the heating-up process attached.

Images attached to this comment