Laser Status:
SysStat is good
Front End Power is -0.004753W (should be around 30 W)
HPO Output Power is 73.44W
Front End Watch is GREEN
HPO Watch is RED
PMC:
It has been locked 0 days, 0 hr 1 minutes (should be days/weeks)
Reflected power = 15.37Watts
Transmitted power = 45.19Watts
PowerSum = 60.56Watts.
FSS:
It has been locked for 0 days 0 hr and 0 min (should be days/weeks)
TPD[V] = 2.893V (min 0.9V)
ISS:
The diffracted power is around 1.4% (should be 3-5%)
Last saturation event was 0 days 0 hours and 50 minutes ago (should be days/weeks)
Possible Issues:
Front End Power is Low
ISS diffracted power is Low
LRA out of range, see SYSSTAT.adl
PSL Weekly script still needs to be updated per FRS 11302.