Following Patrick's restarts of the IOCs on h1ecatx1 and h1ecaty1, the DAQ EDCU reconnected to all but 101 channels. All the channels are from the PLC1 systems, approx 70% from h1ecatx1plc1 and 30% from h1ecaty1plc1. I ran a script on h1dc0 which successfully caget'ed all 101 channels, but this did not prompt the EDCU process to reconnect to these channels.
This problem has persisted since 08:20 this morning (over 4 hours) and we think to only way to get these channels into the DAQ will be a DAQ restart. This is scheduled following the ongong charge measurements.
unconnected channel list attached as text file