[Sheila, Haocun, Nutinsee, Terry]
Yesterday we also did a better OMC cavity scan with beams from squeezer.
Compared with last time, we used:
The results are shown in the attachment.
---> This tells us the propagation and loss from Faraday is ~7.34%. (SRM transmission = 32.34%)
In the Cavity:
Nice! A couple of quick questions about:
Where are these values measured? I imagine the input power was measured on SQZT6 (after the SQZ Faraday) and "Power in the cavity" actually means power incident on the OMC (like on AS_C). Is that correct? Is the alignment to the OMC that good because you engaged OMs alignment?
Yes,
input power was measured on SQZT6 (after the SQZ Faraday).
"Power in the cavity" means power incident on the OMC calculating from the scan, including 00 mode and higher modes.
OMC has good alignment because:
1. We are using ASC SERVO DC centering loops to align
2. We have seed power high enough
And there are some more info if you are interested in:
https://git.ligo.org/haocun.yu/lho_squeezing/wikis/Squeezing-Alignment