haocun.yu@LIGO.ORG - posted 10:48, Wednesday 29 August 2018 - last comment - 16:55, Thursday 30 August 2018(43723)
Better OMC Scan with SQZ SEED Beam

[Sheila, Haocun, Nutinsee, Terry]

Yesterday we also did a better OMC cavity scan with beams from squeezer.

 

Compared with last time, we used:

 

The results are shown in the attachment.

     ---> This tells us the propagation and loss from Faraday is ~7.34%. (SRM transmission = 32.34%)

In the Cavity:

 

 

Images attached to this report
Comments related to this report
lisa.barsotti@LIGO.ORG - 13:00, Wednesday 29 August 2018 (43727)

Nice! A couple of quick questions about:

  • Input power from IR PD: 1.24mW
  • Power in the cavity: 0.78mW

Where are these values measured? I imagine the input power was measured on SQZT6 (after the SQZ Faraday) and "Power in the cavity" actually means power incident on the OMC (like on AS_C). Is that correct? Is the alignment to the OMC that good because you engaged OMs alignment?

 

haocun.yu@LIGO.ORG - 16:55, Thursday 30 August 2018 (43759)

Yes,

input power was measured on SQZT6 (after the SQZ Faraday).

"Power in the cavity" means power incident on the OMC calculating from the scan, including 00 mode and higher modes.

OMC has good alignment because:

1. We are using ASC SERVO DC centering loops to align

2. We have seed power high enough

And there are some more info if you are interested in:

https://git.ligo.org/haocun.yu/lho_squeezing/wikis/Squeezing-Alignment