Displaying report 1-1 of 1.
Reports until 15:00, Friday 31 August 2018
H1 TCS (TCS)
georgia.mansell@LIGO.ORG - posted 15:00, Friday 31 August 2018 - last comment - 15:00, Friday 31 August 2018(43762)
ETMY ring heater test results

[Team TCS]

Last night we ran a low power (0.5 W to upper and lower RH stages) long duration (3 hours) ring heater test on ETMY.

The ETM deformation is visible on the Hartmann wavefront sensor, and the change in spherical power is similar to an identical test we did on ITMY two nights ago.

The prism x and prism y values, a measure of the wedging seen by the HWS, indicate that there is some misalignment from the center of the optic, more in yaw than in pitch. This is shown in the middle plot of the first attachment.

The second attachment is a contour plot, comparing a reference time just before the ring heater was turned on, and a "current time" once the prism values had reached a steady state. The hole in the middle might be because I did not reset the frame rate after starting the Hartmann code?

Edit: The hole in the second plot is due to the scaling, set in the plot contour script. I've rerun the script on the h1hwsmsr computer, where TVo has added an autoscale option, the result is attached as the thrird plot.

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 06:09, Friday 31 August 2018 (43767)

Awesome!!! That hole is a bit odd though, I don't see how the frame rate would affect this. Changing the frame rate should only lead to longer/shorter exposures.

Displaying report 1-1 of 1.