cheryl.vorvick@LIGO.ORG - posted 13:47, Tuesday 04 September 2018 (43812)
PSL Status Report - Weekly: FAMIS Task 10372

The report shows that there was work on the PSL this morning, so locked times are short, and this is understood.  The HPO has no power.  The diffracted power at 1.3% is closer to what we plan to run during O3, so looks OK.

Laser Status:
SysStat is good
Front End Power is -0.003792W (should be around 30 W)
HPO Output Power is 0.1943W
Front End Watch is GREEN
HPO Watch is RED

PMC:
It has been locked 0 days, 7 hr 38 minutes (should be days/weeks)
Reflected power = 14.9Watts
Transmitted power = 51.21Watts
PowerSum = 66.1Watts.

FSS:
It has been locked for 0 days 6 hr and 5 min (should be days/weeks)
TPD[V] = 2.944V (min 0.9V)

ISS:
The diffracted power is around 1.3% (should be 3-5%)
Last saturation event was 0 days 2 hours and 11 minutes ago (should be days/weeks)

Possible Issues:
Front End Power is Low
ISS diffracted power is Low
LRA out of range, see SYSSTAT.adl