SEI seismometer mass position check - Monthly: FAMIS Task 7533
There are ten T240s out of range, and some are about 6 times the maximun range of 0.3V.
There is one STS proof masses out of range at 4 times the maximum of 2V.

Averaging Mass Centering channels for 10 [sec] ...
2018-09-04 13:54:14.359578

There are 10 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 1 DOF Z/W = -0.623 [V]
ETMX T240 2 DOF X/U = -1.175 [V]
ETMX T240 2 DOF Y/V = -1.114 [V]
ETMX T240 2 DOF Z/W = -0.52 [V]
ETMX T240 3 DOF X/U = -0.339 [V]
ETMX T240 3 DOF Y/V = 0.405 [V]
ITMX T240 1 DOF X/U = -1.244 [V]
ITMX T240 3 DOF X/U = -1.402 [V]
ITMY T240 3 DOF X/U = -1.12 [V]
ITMY T240 3 DOF Z/W = -1.949 [V]

All other proof masses are within range ( < 0.3 [V] ):
ETMX T240 1 DOF X/U = 0.164 [V]
ETMX T240 1 DOF Y/V = -0.273 [V]
ETMX T240 3 DOF Z/W = -0.08 [V]
ETMY T240 1 DOF X/U = -0.185 [V]
ETMY T240 1 DOF Y/V = 0.226 [V]
ETMY T240 1 DOF Z/W = -0.189 [V]
ETMY T240 2 DOF X/U = -0.06 [V]
ETMY T240 2 DOF Y/V = -0.173 [V]
ETMY T240 2 DOF Z/W = -0.088 [V]
ETMY T240 3 DOF X/U = -0.218 [V]
ETMY T240 3 DOF Y/V = -0.203 [V]
ETMY T240 3 DOF Z/W = 0.047 [V]
ITMX T240 1 DOF Y/V = 0.217 [V]
ITMX T240 1 DOF Z/W = 0.099 [V]
ITMX T240 2 DOF X/U = 0.035 [V]
ITMX T240 2 DOF Y/V = 0.158 [V]
ITMX T240 2 DOF Z/W = 0.211 [V]
ITMX T240 3 DOF Y/V = 0.122 [V]
ITMX T240 3 DOF Z/W = 0.045 [V]
ITMY T240 1 DOF X/U = -0.126 [V]
ITMY T240 1 DOF Y/V = -0.105 [V]
ITMY T240 1 DOF Z/W = -0.23 [V]
ITMY T240 2 DOF X/U = 0.033 [V]
ITMY T240 2 DOF Y/V = 0.012 [V]
ITMY T240 2 DOF Z/W = -0.051 [V]
ITMY T240 3 DOF Y/V = 0.007 [V]
BS T240 1 DOF X/U = -0.037 [V]
BS T240 1 DOF Y/V = -0.123 [V]
BS T240 1 DOF Z/W = 0.205 [V]
BS T240 2 DOF X/U = 0.0 [V]
BS T240 2 DOF Y/V = 0.197 [V]
BS T240 2 DOF Z/W = -0.102 [V]
BS T240 3 DOF X/U = 0.013 [V]
BS T240 3 DOF Y/V = -0.191 [V]
BS T240 3 DOF Z/W = -0.221 [V]
Assessment complete.

Averaging Mass Centering channels for 10 [sec] ...

2018-09-04 14:01:15.356812
There are 1 STS proof masses out of range ( > 2.0 [V] )!
STS A DOF X/U = -8.095 [V]

All other proof masses are within range ( < 2.0 [V] ):
STS A DOF Y/V = -0.856 [V]
STS A DOF Z/W = -0.389 [V]
STS B DOF X/U = 0.391 [V]
STS B DOF Y/V = 0.343 [V]
STS B DOF Z/W = -0.422 [V]
STS C DOF X/U = 0.842 [V]
STS C DOF Y/V = 0.922 [V]
STS C DOF Z/W = -0.55 [V]
STS EX DOF X/U = -0.128 [V]
STS EX DOF Y/V = 0.439 [V]
STS EX DOF Z/W = 0.098 [V]
STS EY DOF X/U = 0.094 [V]
STS EY DOF Y/V = -0.035 [V]
STS EY DOF Z/W = 0.51 [V]

Assessment complete.