MICH ASC OLTF measurements

Gabriele, Hang

We took several MICH ASC OLTF at different stages. Please see the attached plot.

First we measured the OLTFs when the MICH was locked using AS_B_45_Q signals (corresponding to GRD states from ENGAGE_DRMI_ASC to DRMI_LOCKED_CHECK_ASC). They correspond to traces REF3 (pink) for P and REF1 (cyan) for Y. The gains were low so we modified the input matrix elements from -50 to -100 to increase the UGF to ~ 1 Hz with 50 deg phase margin.

The red (REF5) and blue (REF4) traces were measured at DHARD_WFS state. At this stage BS was locked using the AS_A_36_Q. The loop gains seem fine there. Although it seemed that in the Y loop there was a feature at 2.2 Hz in both stages (both AS45 and AS36 signal). Its original yet needs to be tracked down.

The measurement template is now in

/ligo/svncommon/IscSVN/iscmodeling/trunk/ALIGOH1/ASC_loops/Measurements/MICH_writeable/MICH_BroadBand.xml

The yaw excitation (channel 3) seems to be good enough to resolve the entire band of interest. For P we used two excitations. Channel 1 (which leads to the attached plot) focuses on the high freq >0.8 Hz part and Channel 0 focused on resolving the features around the main sus resonance.

 