I looked around at signals on the PSL and in the IO path, and found two things. The first, which is probably unremarkable, is that when the RefCav Trans PD went to 0, the ISS QPD, and IO GigiE cameras don't see a change in the beam alignment. The second is that when the PMC heater power changed, going from 30 to 60, that change steered the beams coming out of the PMC, as seen on the ISS QPD, and IO GigiE cameras.
Many plots, but the first and last give the most pertinent information.