J. Kissel Noticed the calibration line spectrum on the main sensitivity wall FOM (nuc3) showed no calibration lines. Opened PCAL Y overview screen and found the OFS railed at ~7.5V (as quickly seen by the trend on the MEDM overview screen; see attached). Trended the OFS PD channel (H1:CAL-PCALY_OFS_PD_OUT16), and found that it railed at Sep 11 2018 @ ~18:00 UTC or at ~11a PDT. This is likely a result of the timing glitch at EY (referenced in LHO aLOGs 43932, 43928). In order to fix, I merely needed to toggle the Loop Enable Switch, H1:CAL-PCALY_OPTICALFOLLOWERSERVOENABLE. Also -- we (Travis and I) didn't store the new amplitude settings for the calibration lines in the SDF system. Upon restart I found calibration line heights at their O2 values. I've also re-changed the line heights to their temporarily higher values while we measure the optical gain with our current, not-yet-amazing, sensitivity. For the record, Former New Change 36.7 OSC1 500 1000 x2 331.9 OSC2 5000 20000 x4 1083.7 OSC3 15000 5000 /3 7.93 OSC4 5000 5000 no change See 2nd and 3rd attached files confirming the change and the capture into SDF.