MICH_DARK_ALIGN works again

We reduced the ASC loops gain by a factor of 4 in the MICH_DARK_ALIGN state in ALIGN_IFO guardian and now it seems to be working again. We should be able to use it for again for doing simple mich contrast defect measurements.

Comments related to this report
gabriele.vajente@LIGO.ORG - 08:39, Thursday 13 September 2018 (43986)

MICH_DARK_ALIGN was not working this morning.

The reason was that the PREP_MICH_DARK_ALIGN state will set the sensing matrix and engage the input of the ASC-MICH filter banks, with zero gain. But FM3 is almost an integrator, with a pole at 0.03 Hz. Therefore any input offset in the error signal was running into this low frequency pole, accumulating history.

I changed the way the ASC-MICH loops are engaged and now MICH_DARK_ALIGN works:

  1. PREP_MICH_DARK_ALIGN will set the sensing matrix, but keep the input switch off for ASC-MICH_*
  2. MICH_DARK_ALIGN will switch on the input and ramp in the gain
  3. Then MICH_DARK_ALIGN will switch on a real integrator and engage the BS M1 stage