Reports until 10:04, Tuesday 09 October 2018
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 10:04, Tuesday 09 October 2018 - last comment - 16:52, Tuesday 09 October 2018(44429)
Reference cavity alignment tune up
Tuned up the alignment into the reference cavity.  The transmission improved from ~2.2V to ~3V.


    Removed the DC block from the AOM (see attached photo for the aftermath).

    Measured the transfer function with the current gain slider settings.  The UGF was ~452 kHz
with about 60 degrees of phase margin.  Two measurements ar attached, one out to 2 MHz and the
other out to 10 MHz.  No big changes were observed.
Images attached to this report
Comments related to this report
peter.king@LIGO.ORG - 16:52, Tuesday 09 October 2018 (44449)
A quick note in passing, the cavity visibility is ~84% (1 - 80/500).  The output of the
locking photodiode was ~500 mV unlocked and ~80 mV locked.