Updated soft offsets

IFO locked at 10 W, moved CSOFT and DSOFT to maximize the recycling gain. Moved also SRM and SR2 to keep good sideband buildups. New offsets in the TRANS QPD:

caput H1:ASC-X_TR_A_YAW_OFFSET -0.237102
caput H1:ASC-X_TR_B_YAW_OFFSET -0.133515
caput H1:ASC-Y_TR_A_YAW_OFFSET +0.279309
caput H1:ASC-Y_TR_B_YAW_OFFSET +0.292131
caput H1:ASC-X_TR_A_PIT_OFFSET -0.019295
caput H1:ASC-X_TR_B_PIT_OFFSET -0.022531
caput H1:ASC-Y_TR_A_PIT_OFFSET +0.061096
caput H1:ASC-Y_TR_B_PIT_OFFSET +0.149773

 