Reports until 15:40, Tuesday 16 October 2012
LHO VE
kyle.ryan@LIGO.ORG - posted 15:40, Tuesday 16 October 2012 (4451)
Changed "Step" settings on IP11 and IP9
Noticed IP11 and IP9 were still outputting 5000V even though their surrounding pressures were < 1 x 10-8 torr -> Made slight corrective adjustments to the pump current values at which point these pumps switch from 5000V to 3000V outputs.  As found, they were programmed to change at pump currents of 5 x 10-5 amps.  This value of current corresponds to a pressure which is lower than 1 x 10-8 torr which is the point they should be switching to result in the most efficient pump speed.  The new value is 1.5 x 10-4 amps.  