Reports until 16:03, Monday 05 November 2018
H1 PSL (PSL)
cheryl.vorvick@LIGO.ORG - posted 16:03, Monday 05 November 2018 (45027)
PSL weekly Report FAMIS: 10985

Laser Status:
Front End Power is 33.26W (should be around 30 W)
70W Output Power is 75.6W
Front End Watch is GREEN
70W Watch is GREEN

PMC:
It has been locked 0 days, 6 hr 29 minutes (should be daysweeks)
Reflected power = 14.87Watts
Transmitted power = 43.86Watts
PowerSum = 58.73Watts.

FSS:
It has been locked for 0 days 3 hr and 17 min (should be dys/weeks)
TPD[V] = 2.939V (min 0.9V)

ISS:
The diffracted power is around 2.2%
Last saturation event was 0 days 3 hours and 17 minutes ag (should be days/weeks)


Possible Issues: