Great!
Pic 1: Two laser modes are visible when scanning the SHG.
Pic 2: After biasing the temperature actuator by ~3GHz, we are back to single mode operations.
Nutsinee Daniel
Increased the diode current by ~0.1A to 2.12A and adjusted the temperature to bring the beat note back. Check within ±400MHz and found no mode hope. Also, adjusted the alignment of the beat note path. It shows -4dBm when it reads 158MHz.