Displaying report 1-1 of 1.
Reports until 09:06, Thursday 08 November 2018
H1 AOS
robert.schofield@LIGO.ORG - posted 09:06, Thursday 08 November 2018 (45118)
No evidence of backscattering noise from HWS table

I mounted an accelerometer and shaker on the HWS table by HAM4, and found no evidence of noise in DARM when I increased the table motion by a couple of orders of magnitude between about 10 and a few hundred Hz (see figure).

Images attached to this report
Displaying report 1-1 of 1.