Reports until 08:51, Monday 19 November 2018
H1 ISC (CDS, INS, ISC)
keita.kawabe@LIGO.ORG - posted 08:51, Monday 19 November 2018 - last comment - 09:48, Monday 19 November 2018(45380)
New LSC-REFL_B RF check (Georgia, Richard, Keita)

We connected the 5-way coax to TNC adapter that Fil has made to the chamber. 5-way coax cable assignment is in D1300466.

cable 1 cable 2 cable 3 cable 4 cable 5
Single ended DC Test in Test out RF LOW (9MHz) RF HI (45MHz)

We measured the TF from cable 2 to cable 4 while cable 5 was terminated (attachment 1), then from cable 2 to cable 4 while cable 4 was terminated (attachment 2) and both made sense.

The problem of this configuration is that shells of TNCs are touching with each other (attachment 3, note that cable 2 and cable 5 are terminated).  We put cable 1, 2, 3 and 5 inside different fingers in a glove to insulate things from each other (attachment 4).

For the moment 9MHz connector is exposed but it's not touching any surrounding metals.

For a short while we locked MC with 200mW and measured the coherence between REFL_A and REFL_B 9MHz, and it also made sense (see Georgia's comments).

Images attached to this report
Comments related to this report
richard.mccarthy@LIGO.ORG - 09:46, Monday 19 November 2018 (45383)

After ground loop checks we will isolate these with heat shrink.

georgia.mansell@LIGO.ORG - 09:48, Monday 19 November 2018 (45384)

I'm attaching the power spectra of, and coherences between, LSC-REFL_A_RF9_I, LSC-REFL_A_RF9_Q, LSC-REFL_B_RF9_I, and LSC-REFL_B_RF9_Q, with and without light on the photodiodes.

Blue, red, green, and brown traces are with the IMC offline.

In the top plot cyan and pink are REFL_A_RF9 with the IMC locked with 200 mW, orange and black are REFL_B_RF9. Signal is seen on both photodiodes, there is a gain difference between them.

Bottom left plot shows REFL_B_RF9_I and Q coherence with REFL_A_RF9_I; bottom right plot shows REFL_B_RF9_I and Q coherence with REFL_A_RF9_Q. Higher coherence is seen with REFL_A_RF9_I, not surprising since REFL_B has not been phased.

Images attached to this comment