Attached is the trend data plot for the flow rates in the power meter circuit, the front end laser, and the 70 W amplifier.
Lots of glitches, not sure of the reason why. Especially when compared to the 70 W amplifier cooling circuit.
The diode chiller is indicating an error message of "F3-error", which is something related to the hot gas valve. That
might be an indication that the diode chiller needs servicing, I don't know. The diode chiller temperature over the same
period of time looks pretty constant at 20 +/- 0.2 degC. The diode chiller flow looks pretty constant over the same time
interval as well.
The flow rate from the crystal chiller is noticeably lower upon restart. 14 lpm cf. 15.7 lpm.
After restarting the laser, the power meter flow somewhat regularly flashes red indicating that the flow rate has
dropped to 0.5 lpm or lower.