Squeezer recovery work after pumping down:
- Squeezer laser was back on
- SHG was locked
- Green fiber polarization adjustment: Very good. Reached to only 0.00~0.01mW on SHG Fiber Rejected by adjusting the HWP/QWP on ISCT6 and stable.
- PDs were re-aligned, including OPO Trans, OPO REFL, OPO REFL Rejected and OPO IR.
- SEED input diode re-calibrated
A quick cavity scan with green beam, as attached below.
Calculation:
- 00 mode: 89%
- Misalignment: 5.8%
- Mis-mode matching: 3.8% (Mode matching efficiency = 96%)
- A higher mode: 1.4%
This is pretty much better than before, and consistent with the scan in-air.