H1 IOO (IOO)
cheryl.vorvick@LIGO.ORG - posted 08:34, Tuesday 27 November 2018 (45517)
power measure ments and beam scan on the IO path on the PSL

I use 200W power meter to measure powers in the IO path on the PSL.  Adjusting for the changes in PMC Trans , the measurements are:

         
measured PMC TRANS PMC Locked 49.500 W
calculated ALS beam 0.025 1.238 W
calculated EOM Incident:   48.263 W
measured EOM beam PMC Locked 47.300 W
calculated power on beam dump   0.963 W
         
         
         
measured EOM beam PMC Locked 47.300 W
calculated EOM rejected 0.005 0.237 W
calculated EOM pick off 0.0003 0.014 W
calculated EOM main beam   47.049 W
measured Bottom Periscope 49W Req. 46.960 W
calculated lost due to other sources   0.089 W

 

Attached: BEAM SCAN of beam reflected beam from IO_AB_W1, color changed on 2nd and 4th image

 
Images attached to this report