Following on from Keita's alog entry from yesterday this morning
I looked at the mixer monitor signal. The MEDM screen screen indicates that the mixer output is typically ~0.5 V. With a digital
multimeter I measured 1.2 mV when the MEDM screen indicated that the voltage was ~0.7 V. An oscilloscope trace of the signal
is shown in tp3_1.png.
CG20FG9.tif shows the measured open loop transfer function from this morning, with the marker placed at the UGF.
The transfer function does not appear as smooth as ones I can recall from before. Phase margin looks okay.
Whilst I was in the enclosure, I tweaked the alignment to the reference cavity a little and re-centered the beam
onto the locking photodiode. The FSS still has some trouble locking even when the transmission voltage indicated on
the MEDM screen exceeds the resonant threshold. I do not know why this would be, although sometimes either the runtime
model and/or guardian takes over the common gain slider and moves it at what I think is an inappropriate time - like
when there is no light flashing in the reference cavity, for example.