J. Kissel
ECR E1800246
IIET 11305
I've (nearly*) completed several MEDM screens that are reflecting the new upgraded time-dependent correction factor calculation in the fron-end calibration pipeline, in support of splitting the calculation of kappa_pu into kappa_p and kappa_u, plus the new supporting calibration line, as described in T1700106. The changes were installed into the CAL-CS front-end model moons ago (see LHO aLOG 44459, and G1801594), and thus I'm just completing the effort started in LHO aLOG 44752 to develop a user interface.
This system doesn't work yet (we also don't have an IFO back yet to test it), but now that the use interface is complete enough to debug the parameters and settings and we can finally move forward.
*There're a few bug fixes and a bit of aesthetic cleanup let to be done, but we'll get to that in due time.
These new screens live in the following corner userapps repo:
/opt/rtcds/userapps/release/cal/common/medm/
with names akin to the file names of the respective screenshots attached below.
Also in due of time, I'll update G1801594 to reflect the new changes.
WP 7968
Over the course of developing these screens, I'd identified a few tweaks, extra EPICs records, and bug-fixes to the front code along the way, so Dave graciously agreed to support a couple of h1calcs model, and subsequent DAQ, restarts yesterday evening, a. la. LHO aLOGs 45539 and 45543.
These model changes have been committed to
/opt/rtcds/userapps/release/cal/common/models/
as a part of the clean-up.