Sheila, Haocun, Daniel, Terry, Nutsinee
Green power was drifting slowly when we were taking the measurement so could be better. OPO locking crapped out just after we took the data so we didn't have more tonight. Haocun is working on the loss estimate. More recovery alog to follow.

Nice recovery, congratulations on your 5 dBs!
Let's wait for Haocun's analysis, but a quick check suggests that the sqz/anti-sqz numbers are not consistent with the NLG of 7. Getting 8.5dB of anti-sqz with NLG=7 requires huge loss, which would make impossible to achieve 5 dB of sqz.
I agree. Sadly we measured anti squeezing first when the input power to the coupler was ~18-19mW. By the time we get to measuring the squeezing the power dropped by at least a mW. More data is needed.