S. Karki, J. Kissel As sudarshan pointed out last week (see LHO aLOG 45587), there were some bugs / inconsistencies between the recent front-end implementation of the explicit calculation of the PCAL optical efficiencies and force coefficients (calculations described in the appendix of T1800046; implemented in LHO aLOG 44418). I've fixed the model (new screen shots attached), and have restarted the front-end code for both h1calex and h1caley, both of which use the library part I changed, /opt/rtcds/userapps/release/cal/common/models/PCAL_MASTER.mdl which has been committed to the repo. The updated screen is also attached, and committed to /opt/rtcds/userapps/release/cal/common/medm/PCAL_END_FORCE_COEFF.adl