Displaying report 1-1 of 1.
Reports until 20:09, Tuesday 11 December 2018
H1 ISC
sheila.dwyer@LIGO.ORG - posted 20:09, Tuesday 11 December 2018 (45856)
dark noise

While we are having some wind I unlocked and misaligned the mode cleaner and measured the OMC DCPD dark noise with different whitening filters on.  The configuration we have been using is one stage of whitening.  It looks like we don't have much to win in terms of dark noise by turning on the additional stages.  

 

Images attached to this report
Non-image files attached to this report