Reports until 23:15, Thursday 27 December 2018
H1 SQZ (SQZ)
nutsinee.kijbunchoo@LIGO.ORG - posted 23:15, Thursday 27 December 2018 - last comment - 14:14, Monday 31 December 2018(46181)
SQZ pump laser current moved further away from multimode

Previously the pump laser has been drifting in and out of multimode. Today I came in and found it constantly running multimode (an obvious symptom being low green production out of the SHG). So I put the current half way between where the multimode starts and clamp current (and set the temperature to keep PSL and pump beat note at 158 MHz beat note at zero MHz offset on the TTFSS temp servo). The multimode was monitored with the SHG GR PD (Thorlabs) while scanning the SHG cavity. Below is a few data point regarding this particular shitty laser. 

 

  Current (A) Temperature (deg C)
Before (constantly running multimode) 2.134 29.69
Now (no more multimode) 2.193 29.65
Multimode starts (approaching from clamp) 2.137 29.68
Clamp 2.250 29.59
Mode hop (beat note disappeared) 2.050 29.75

These numbers were read of the controller box LCD screen. The current readback on the medm screen aren't the same. 

 

Keeping the temperature the same I tweaked the current to figure out MHz/A calibration

Current (A) Beat note (MHz)
2.193 164.2
2.136 30.2
2.250 308.2
2.222 235.2

This set of data yields 2.5MHz/mA (+- 0.1MHz). 

 

Then I keep the current the same, and tweaked the temperature

Temperature (C) Beat note (MHz)
29.65 164.8
29.63 89.1
29.62 48.2
29.69 290.8
29.72 361.1

This yields 34.21MHz/0.01degC (+-6.6MHz) 

 

If I calculated things correctly here this means we are +70 MHz away from multimode at the moment. The temperature would only have to drift .02 deg C to go back to multimode. Could there be a region beyond mode hop that works?

Comments related to this report
sheila.dwyer@LIGO.ORG - 14:14, Monday 31 December 2018 (46196)

There should be more mode hop free regions at lower currents (so you can get away from the max current).