Reports until 11:31, Thursday 03 January 2019
H1 PSL (PSL)
corey.gray@LIGO.ORG - posted 11:31, Thursday 03 January 2019 (46220)
PSL Weekly Report (FAMIS #10993)

Possible Issues:  NONE to note
Laser Status:
Front End Power is 32.18W (should be around 30 W)
70W Output Power is 71.57W
Front End Watch is GREEN
70W Watch is GREEN

PMC:
It has been locked 0 days, 18 hr 8 minutes (should be days/weeks)
Reflected power = 10.2Watts
Transmitted power = 55.36Watts
PowerSum = 65.57Watts.

FSS:
It has been locked for 0 days 0 hr and 11 min (should be days/weeks)
TPD[V] = 2.529V (min 0.9V)

ISS:
The diffracted power is around 1.7%
Last saturation event was 0 days 0 hours and 12 minutes ago (should be days/weeks)