Locking today

 

 

 

  • In an intermediate locking attempt another earthquake from Japan hit as we were engaging ASC for full IFO. Hard loops oscillated, all TM PUMs tripped, violin modes got rung up. Had a problem with ETMY violins - mode 1 and mode 6 are very close together in frequency, and one of them was getting rung up with the guardian damping settings. Turned off the damping feedback to mode 6, reduced the gain on the feedback to mode1, this did the trick.

 

  • Dolphin crash (05:31:05 UTC) caused ETMY trip while in ADJUST_POWER, lock lost immediately.
    • After Dave restarted all the EY models I untripped the watchdogs. I was able to get up to FIND_IR again at one point, but ALS was being flakey so I gave up locking for the night.
Comments related to this report
peter.fritschel@LIGO.ORG - 06:11, Monday 07 January 2019 (46252)

The L1 ASC dither frequencies are all in the range of 6-9 Hz. Why were the H1 dither frequencies set so much higher in the first place?