The newly rebuilt 2500 l/s ion pump (at X-mid) is ready for installation tomorrow. Gerardo and I had already tested it on the VEA floor last month. I vented it with UHP N2 this afternoon and removed the fasteners from its 16.5" blank flange. I climbed around on the recently erected scaffolding (thanks Chris S., Bubba G.) -seams sturdy and is tagged Ready for Use. I also positioned pump carts and gas bottles etc...
After energizing the MTP controller to charge the levitation batteries and noticed that the long-idle Turbo (stopped) was 2 Torr and the foreline was 780 Torr. There is a known leak in the foreline. As the manual isolation valve at the Turbo Header is closed, this should have remained at rough vacuum. I removed all of the KF fittings that I could and blanked-off just past the leak detector connection and pumped out this new minimal volume with the scroll pump -> Though the scroll pump can bring this minimal-fitting foreline down to 3.5 x 10-2 Torr, the foreline pressure rises rapidly once the scroll pump is isolated. I'll investigate more tomorrow. For tonight, we won't run the Turbo.