Displaying report 1-1 of 1.
Reports until 15:42, Wednesday 09 January 2019
H1 PSL
daniel.sigg@LIGO.ORG - posted 15:42, Wednesday 09 January 2019 - last comment - 16:21, Wednesday 09 January 2019(46317)
ISS Second Loop Modifications

Following up on alog 46240, we made the following mods to the ISS Second Loop, S/N S1700062, which is now installed again (PCB D1600298):

To account for the increased analog gain we reduced the VGA gain to 0dB (from +20dB). We increased the REFERENCE_IN_MTRX_1_2 gain to 10 (from 1) and the REFERENCE_IN_MTRX_1_4 gain to 0.5 (from 0.05). The reference offset was also adjusted to 537 (from -35.14). At first glance, the output offset fluctuations are significantly reduced.

Non-image files attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 16:21, Wednesday 09 January 2019 (46322)

H1:PSL-ISS_SECONDLOOP_REFERENCE_IN_MTRX_1_4 is controlled by IMC_LOCK guardian, so a factor of 10 was put into that (i.e. 0.5 from 0.05 for AC coupled, and 50 from 5 for DC coupled 30W).

Also we noticed that 10Hz pole on/off indicator for 3rd loop and slow offset path on the front panel was not working (it never turned on), but the readback was working and the pole seemed to be switching.

Displaying report 1-1 of 1.