Reports until 15:30, Friday 11 January 2019
H1 SQZ (SQZ)
nutsinee.kijbunchoo@LIGO.ORG - posted 15:30, Friday 11 January 2019 (46376)
SQZ laser temperature knob found locked and away from nominal

The green power out of the SHG became lower around noon and I noticed that the temperature knob on the laser box has been locked. I left it unlocked the last time I adjusted it because the knob was too sensitive. By locking the knob the temperature was changed by .05 degree C. That was enough to put us in the multimode region. Again, 

PLEASE DO NOT TOUCH ANY KNOBS ON THE SQZ LASER CONTROLLER BOX. 

The setting has been reversed back and temperature knob was left unlocked.