Reports until 23:18, Tuesday 15 January 2019
H1 ISC
sheila.dwyer@LIGO.ORG - posted 23:18, Tuesday 15 January 2019 - last comment - 17:32, Wednesday 16 January 2019(46459)
ASC slow instabilities much better, changes not in guardian

Sheila, TVo, Georgia, Craig, Danny

This current ASC situation seems better than what we've had the last several days, so we probably want to make some of these changes permanent, but we don't want to make any changes in the guardian since ASC engagement has been tricky and we may not have a chance to test things. I think that turning off PRC1Y, and increasing the CHARDY gain from 3 to 6 are definitely improvements.  The change to the CSOFT input matrix I'm not so sure about. 

Comments related to this report
jenne.driggers@LIGO.ORG - 17:32, Wednesday 16 January 2019 (46486)

I've put the CHARD_Y gain increase and the PRC1_Y loop-off at the end of ENGAGE_SOFT_LOOPS.  I have not changed the SOFT input matrix.

In addition, since we haven't been using the DSOFT_P loop lately, I've told the guardian to stop turning on the DSOFT_P radiation pressure compensation.  Now, when guardian is turning on the RPC stuff, if the DSOFT_P gain is zero, then it skips over DSOFT.