Displaying report 1-1 of 1.
Reports until 00:55, Wednesday 16 January 2019
H1 ISC (ISC)
georgia.mansell@LIGO.ORG - posted 00:55, Wednesday 16 January 2019 - last comment - 16:25, Friday 18 January 2019(46461)
SRCL feedforward tuning

Craig, Georgia, Danny, Sheila

We have re-tuned the SRCL feedforward for 30 W interferometer configuration. The updated feedforward reduced SRCL coupling to DARM from 60Hz to 500Hz (driven measurement shown in final attachment).

We measured the transfer function from SRCL to DARM without feedforward, and from the SRCL-FF path to DARM, using templates found here:

/opt/rtcds/userapps/release/lsc/h1/scripts/feedforward/

We then used Craig's ipython notebook to fit these transfer functions and calculate the feedforward filtering required to cancel SRCL noise in DARM. The notebook is found here:

/ligo/home/craig.cahillane/Git/Feedforward/SRCL2DARMfeedforward.ipynb

This notebook uses iirrational to fit the transfer functions. We added some human tweaking to the fit to minimise the resonant peak at ~8Hz. First attachment shows the data (blue), iirrational's fit (orange), and the human-modified fit that was implemented in the feedforward filter (green). Note that the fit to the phase of the TF is not good below 60 Hz, so perhaps there is room for iterative improvement on these filters.

The second attachment shows the old SRCL-FF filter (Nov7) compared to our new filter (Jan15).

The final attachment shows the SRCL to DARM coupling reduction: yellow is DARM (and SRCL) with no excitation, blue is with the old feedforward, purple is the new feedforward.

Images attached to this report
Comments related to this report
daniel.vander-hyde@LIGO.ORG - 16:25, Friday 18 January 2019 (46535)

Attached is a similar measurement but on 1/18/2019. Not exactly the same excitation but close enough to show that with the same feed forward filter the SRCL noise is higher (between 60 and 450 Hz) than it was when we took this measurement a couple of days ago.

Images attached to this comment