Both of these QPDs acquired an NSUM channel. The medm screens reached from the IMC overview have been updated to also include the whitening.
The calibration is now according to alog 46482. The segments are in mW, whereas the new NSUM channels is in W incident to the mirror.
The IMC trigger had to be adjusted to account for the change in the MC2 TRANS SUM.
We also updated the ASC-AS_C QPD calibration. The previous calibration was 18dB too low, because the whitening gain was changed to 18dB from 36dB. We added the mW calibration into the segments and added a calibration relative to the incident mirror power (power towards HAM6) into the NSUM filter module.