FSS measurements (continued)
Following on from yesterday's attempt.  I made transfer function and spectra measurements with the input modecleaner locked and
unlocked.  In summary, there wasn't too much difference in the noise spectra above 100 kHz with the input modecleaner locked
or unlocked.

    Attached are transfer function measurements when the input mocdecleaner was unlocked and locked respectively (IMCULKD1.tif,
IMCLKD1.tif).  The noise spectra measurements are IMCULN1.tif and IMCLN1.tif respectively.  The mixer calibration was measured
to be 1.93/77.4 V/kHz.

    More to follow ...
Images attached to this report
Comments related to this report
peter.king@LIGO.ORG - 11:44, Thursday 17 January 2019 (46503)
mxr_imc.txt: FSS mixer monitor signal from 100 Hz to 2 MHz, measurement taken with 4395A and active probe
             plotted as hfmxr.png

c1.txt: mixer monitor signal at the carrier, zoomed in
        plotted as carrier.png

MXR[1-4].txt and MXRIMCL[1-4].txt: FSS mixer monitor signal from 10 Hz to 100 kHz, measurement taken on
                                   SR7855, plotted as mxr.png
Images attached to this comment
Non-image files attached to this comment
craig.cahillane@LIGO.ORG - 17:55, Thursday 24 January 2019 (46631)
I stitched and calibrated Peter's FSS mixer data with IMC locked from the SR785 and Agilent together, multiplying the Agilent spectrum by a factor of 25 to get them to match up.
Images attached to this comment