Displaying report 1-1 of 1.
Reports until 09:12, Monday 28 January 2019
H1 PSL
jim.warner@LIGO.ORG - posted 09:12, Monday 28 January 2019 (46670)
PSL Weekly

Laser Status:
Front End Power is 32.25W (should be around 30 W)
70W Output Power is 71.1W
Front End Watch is GREEN
70W Watch is GREEN

PMC:
It has been locked 5 days, 18 hr 17 minutes (should be days/weeks)
Reflected power = 10.65Watts
Transmitted power = 54.35Watts
PowerSum = 64.99Watts.

FSS:
It has been locked for 0 days 2 hr and 43 min (should be days/weeks)
TPD[V] = 3.271V (min 0.9V)

ISS:
The diffracted power is around 2.4%
Last saturation event was 0 days 4 hours and 28 minutes ago (should be days/weeks)


Possible Issues:
 