H1 ISC
marie.kasprzack@LIGO.ORG - posted 13:50, Thursday 07 February 2019 (46800)
Study and variations of DSOFT P loop

Over the last days, we did several measurements of the DSOFT and DHARD loops to investigate the oscillations that lead to loose lock. In particular 0.47 Hz and 1.1 Hz oscillations are seen frequently.

With the angular controls offloaded to M0, we first checked the cross-over frequency between M0 and L1. We drove the two stages through the hierarchical path and measured the angular signals with the oplevs. The model is in very good agreement with the measurements and we can extrapolate that the cross-over frequency is slightly above 0.1 Hz (see figure 1). We also check the angular LOCK filters in all test masses are the same, which is mostly true, expect for a gain difference in M0 ETMX P (see figure 2 and 3).

DSOFT P: The loop was measured at 20W without the Radiation Pressure Compensation (RPC) and without the RESG filter. We observed an unexpected phase shift around 0.46 Hz.

I would suggest to decrease the M0 gain by a factor 3 to decrease the crossover, and remeasure the loop to see if this removes the instability.  I would also recommend to increase the gain if this is compatible with the noise coupling in the IFO.

 

Images attached to this report