Notes on the procedure/measurements:
- Step 1c: (HV values) For ETMy it's HVMON values were low (i.e. -3). Think I noticed this last time I did this on 12/5/18
- Step e&f: The ETMy has a much different pitch value than a few months ago. That's fine, but I did notice Oplev pitch was pretty high over -30. I just went ahead and zero-ed this
- General Prep Note: ETMx & ETMy look different. ETMx ESD Coil outputs all have values, on order of 50k (vs. ETMy whose similar channels are all zero). I don't look at these screens enough, but maybe this is due to how we use the ESD on ETMx and how ETMy continues to have a bad ESD (i.e. the LR).
- For ETMx measurement (see attached), the 1st data file had errors, so I only had (4) good measurements.
- For ETMy measurement (see attached), the 1st & 4th data file had errors, , but als had (4) decent measurements (with L
- Long Tred Plot (see attached) note:
- For ETMx made stop time: 2019,02,12,23,00,0
- Used usual Long_Trend.m script.
- ETMx ESD Charge Eff Bias Trend showed only the Mean data point for my measurement.
- ETMX ESD Actuatioon Strength Trend showed all the data points at 0.
- For ETMY made stop time: 2019,02,12,23,00,0
- Used Long_Trend_ETMY_LHO.m script.
- ETMy ESD Charge Eff Bias Trend showed only the Mean data point for my measurement.
- ETMY ESD Actuatioon Strength Trend showed all the data points at 0.
- Let both charge scripts run to completion. Returned pit/yaw sliders for both optics. Returned ETMx Cal Line amplitude to 0.55 (from 0 from measurement).
Here are the original Slider Values (which I returned to after the measurement):
ETMx:
ETMy: