jameson.rollins@LIGO.ORG - posted 01:19, Saturday 09 February 2019 - last comment - 12:41, Friday 15 February 2019(46883)
Exploring 9MHz RIN situation

[Rana, Jamie]

We spent the evening trying to get more measurements of the 9MHz RIN situation, basically redoing what was done in LHO log 46586.  We went to ITMY single-bounce and locked OMC on carrier, then went to PRMI and locked OMC on 9MHz.  I'll post plots tomorrow.

FYI the OMC and PRMI configurations might be a little bit funky.  We didn't get to reset everything yet to get back to low noise, just in case any one tries to brave the snow to get back out here.

Comments related to this report
jameson.rollins@LIGO.ORG - 14:19, Monday 11 February 2019 (46889)

Friday night Rana and I re-measured the relative intensity noise (RIN) in the carrier and 9MHz sideband at the OMC DCPD.  We believe we got a lower noise measurement (first attachment) than what Craig and Koji measured previously (LHO 46586) by:

  • increasing the OMC length dither amplitude by x300 (the dither amplitude had been tiny, we should probably address this permanently)
  • using PRMI lock for the 9MHz measurement

For the carrier measurement used single bounce off of ITMX with 25W input, and for the 9MHz used PRMI (where the 9MHz sideband is largest) with 35W input.

We don't see any excess RIN in the 9MHz above 300 Hz.

1233733816 carrier measurement
1233738050 9MHz measurement
1233738400 dark noise measurement
Images attached to this comment
gabriele.vajente@LIGO.ORG - 20:24, Monday 11 February 2019 (46890)

1) the carrier RIN in your plot at 30 Hz is in the 1e-7 region. Do we believe this excess above the ISS second loop measurement and above the noise floor? Or is it OMC length noise?

I’m asking because a RIN of 2e-7 is what we need to be limiting DARM, see https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=46759

If we believe the carrier single bounce RIN measured by Jamie and Rana, and use the radiation pressure intensity noise coupling (based on 8% power imbalance and matching Craig's measurement in 46817), we get something pretty close to the DARM noise:

 

2) same question for the excess SB RIN? Is that PRMI control noise?

Images attached to this comment
daniel.sigg@LIGO.ORG - 10:57, Tuesday 12 February 2019 (46900)

For 35W input power, the ISS inner PD shows 27.3mA, whereas the outer shows 29.7mA.

jameson.rollins@LIGO.ORG - 15:10, Wednesday 13 February 2019 (46935)

Note I think it's possible that we were looking at the 45MHz sideband in the PRMI lock, rather than the 9MHz sideband.  We will attempt to retake this measurement once the IFO is recovered.

rana.adhikari@LIGO.ORG - 12:41, Friday 15 February 2019 (46964)ISC

it seems that the OMC length lock is not strong enough to measure low RIN - this is why the increase of the OMC length dither decreases the RIN.

Most of the low frequency peaks are due to acoustics - we were using the QPDs for OMC-ASC instead of angular dither.

Displaying report 1-1 of 1.