jameson.rollins@LIGO.ORG - posted 16:55, Wednesday 13 February 2019 - last comment - 17:38, Wednesday 13 February 2019(46937)
PSL tripped

The PSL is offline again.  Seems the chiller tripped again.  XTALTEMP started rising about 5 minutes ago.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 17:38, Wednesday 13 February 2019 (46938)

We went to the chiller, in the upper chiller we see no water in the tube where you fill the water.  The display says Outlet temperature 20.7C setp: 20C, and cycles through error messages: flow senor 1, conductivity too high, level low. 

We saw some water on the floor about 1-2 cups of water.  Some photos that show where the water was are attached.  We wiped all the water up that was outside the chiller, but didn't reach in and around the cables to dry up inside the enclosure.  The gray canvas in the back seems completely dry.

Images attached to this comment