Reports until 18:30, Sunday 24 February 2019
H1 ISC
georgia.mansell@LIGO.ORG - posted 18:30, Sunday 24 February 2019 (47108)
ALS X glitchiness

Craig, Georgia

This afternoon we had some problems locking the X arm in green. We found it would acquire, lock briefly, then lose lock after a few seconds, similar behaviour to what I saw back in January. We spent a while adjusting alignment, tried toggling the noise eater, locking without the WFS, finally Craig adjusted the PDH locked gain. When he set it back to its nominal value (4), a few things happened within a second:

I don't understand why locking would cause the power from one of the laser diodes would change, or why this particular lock worked out. Is it a coincidence and the laser just happened to hop out of a mode-hopping region?

 

Images attached to this report